Light source: 6V 10W, halogen lamp with more than 2000 hours’ life time
Spectrometer: Interference filter
Wavelength: 340nm, 380nm, 405nm, 505nm, 546nm, 578nm, 620nm, and one free position
Flow cell: Permanent quartz flow cell, minimal volume is 32μL, and disposable cuvette is selectable
Test method: End point, Fixed time, Kinetic, Factor, Multi-standards, Bichromatic, Dual wavelengths
Screen: Touch screen
Stray light: ≤0.5%(absorbance≥2.5)
Absorbance range: 0 ~3.000
Absorbance linearity: ±1%
Absorbance repeatability: ±0.001
Absorbance resolution: 0.0001
Precision: ≤1.0%
Carry over: ≤1%
Temperature: 25℃, 30℃, 37℃, ±0.3℃
Aspirationvolume: 200μL~3000μL
Sample volume: 10μL~50μL
Reaction curve: Real time display the reaction curve
Memory: Able to store more than 200 parameters while 55 parameters of them are stablized; Able to save more than15000 test results and auto save the results if the power is off during working
QC: Auto draw the QC figure
Reagent: Open reagent
Display: LCD with high brightness
Input: Key board and touch screen
Printer: Inner thermal printer, support on-line printing
Input and output: RS-232 serial port, free software to connect with LIS